Model 200E Instruction Manual
Getting Started
Perform the zero/span calibration procedure:
M-P CAL
NOX STB=XXX.X PPB
NO = X.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
NOX STB=XXX.X PPB
NO2= X.X
< TST TST >
CAL
SETUP
M-P CAL
NOX STB=XXX.X PPB
NOX= X.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
NOX STB=XXX.X PPB
NO2= X.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL
RANGE = 500.0 PPB NOX = X.X
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL
RANGE = 500.0 PPB
03 =
< TST TST > ENTR CONC
EXIT
Wait until
NOX STB
falls below
0.5 ppb.
This may take several
minutes.
The
SPAN
key now
appears during the
transition from zero to
span.
You may see both keys.
If either the
ZERO
or
SPAN
buttons fail to
appear see Section 11
for troubleshooting tips.
SAMPLE RANGE = 500.0 PPB NOX= X.X
< TST
TST
>
CAL
SETUP
SAMPLE
NOX STB=XXX.X PPB
NO = X.X
< TST TST >
CAL
SETUP
Set the Display to show the
NOX STB
test function.
This function calculates the
stability of the NO/NO
x
measurement
Press
ENTR
to changes the
OFFSET
&
SLOPE
values for both
the NO and NO
x
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
NOX STB
may jump
significantly.
Wait until it falls back
below
0.5 ppb.
This may take several
minutes.
Press
ENTR to
change the
OFFSET
&
SLOPE
values for both
the NO and NO
x
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
Analyzer
continues to cycle
through NO
x
, NO,
and NO
2
measurements .
The Model 200E analyzer is now ready for operation.
3.6.2. Interferences for NO
X
Measurements
The chemiluminescence method for detecting NO
X
is subject to interference from a number
of sources including water vapor (H
2
O), ammonia (NH
3
), sulfur dioxide (SO
2
) and carbon
dioxide (CO
2
) but the Model 200E has been designed to reject most of these interferences.
Section 10.1.5 contains more detailed information on interferences.
044100102 Rev A
27
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...