EPA Protocol Calibration
Model 200E Instruction Manual
Table 8-1: Activity Matrix for EPA Calibration Equipment and Supplies
Equipment/
Supplies
Acceptance Limits
Frequency And Method
Of Measurement
Action If Requirements
Are Not Met
Recorder
Compatible with output signal
of analyzer; min chart width of
150 mm (6 in) is recommended
Check upon receipt
Return equipment to
supplier
Sample line
and manifold
Constructed of PTFE, glass or
stainless steel
Check upon receipt
Return equipment to
supplier
Calibration
equipment
Meets guidelines and Section
2.3.2 of Q. A. Handbook
See Section 2.3.9 of Q.
A. Handbook
Return equipment/
supplies to supplier or
take corrective action
Working
standard NO
cylinder gas
Traceable to NIST-SRM. Meets
limits in traceability protocol for
accuracy and stability.
Section 2.0.7 of Q. A. Handbook
Analyzed against NIST-
SRM; see protocol in
Section 2.0.7, Q.A.
Handbook
Obtain new working
standard and check for
traceability
Recording
forms
Develop standard forms
N/A
Revise forms as
appropriate
Audit
equipment
Cannot be the same as used for
calibration
System must be
checked out against
known standards
Locate problem and
correct or return to
supplier
Table 8-2: Activity Matrix for Calibration Procedure
Equipment/
Supplies
Acceptance Limits
Frequency And Method
Of Measurement
Action If Requirements
Are Not Met
Calibration
gases
Sec. 2.0.7, Q.A. Handbook
Assayed against an
NIST-SRM quarterly,
Sec. 2.0.7
(Q.A. Handbook)
Working gas standard is
unstable, and/or
measurement method is
out of control; take
corrective action such as
obtaining new calibration
gas
Dilution gas
Zero air, free of contaminants;
TAD
2
and Sec. 2.0.7, Subsec.
7.1, Q.A. Handbook
See TAD
2
Return to supplier or take
appropriate action with
generation system
Multi-point
calibration
(GPT)
t
R
< 2 minutes P
R
> 2.75 ppm/min
Use calibration procedure in Q.A
Handbook; also CFR
1
Converter efficiency > 96%
Method: Section 7.2 of
this manual)
CFR Title 40
1
,
Appendix F
Adjust flow conditions
and/or reaction cell
volume to meet
suggested limits
Repeat the calibration
Replace or service the
converter
128
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...