Operating Instructions
Model 200E Instruction Manual
•
Automatic range mode (
AUTO
) gives the analyzer the ability to report data through
a low range and high range by automatically switching into the appropriate range as
the concentrations change.
Only one of the above range modes can be active at any time. To select output range types,
use the following key sequences.
6.5.3. Single Range mode (SNGL)
The default range mode for the analyzer is single range, in which all analog concentration
outputs (usually
A1, A2 and A3
) are set to the same reporting range. This reporting range
can be set to any value between 100 ppb and 20000 ppb. However, the electronic signal
ranges of the analog outputs may still be configured for different values (e.g., 0-10 V and
0-0.1 V).
To select
SNGLE
range mode and to set the upper limit of the range, press:
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP C.3
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP C.3
RANGE MODE:
SNGL
SNGL
DUAL AUTO
ENTR EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP C.3
RANGE MODE:
SNGL
SNGL
DUAL AUTO
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP C.3
RANGE:
500.0 Conc
0 0
5
0 0 .0
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
NOTE
If a value <100.0 or
>20.000 is entered,
the
ENTR
key will
disappear
6.5.4. Independent Range Mode (IND)
The independent range mode allows the concentration outputs
A1, A2
and
A3
to be
configured with a different range value. In
IND
range mode the
RANGE
test function
displayed on the front panel will then be replaced by three separate functions:
•
RANGE1
: Range value for output
A1
(NO
x
), e.g., 0-10 V for 0-1500 ppm NO
X
.
•
RANGE2
: Range value for output
A2
(NO), e.g., 0-10 V for 0-1000 ppm NO.
•
RANGE3
: Range value for output
A3
(NO
2
), e.g., 0-10 V for 0-500 ppm NO
2
.
50
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...