
Operating Instructions
Model 200E Instruction Manual
6.5.5. Auto Range Mode (AUTO)
In AUTO range mode, the analyzer automatically switches the reporting range between two
user-defined ranges (low and high). The same low and high span settings are applied
equally to NO, NO
2
and NO
X
readings. The unit will switch from low range to high range
when either the NO, or NO
X
concentration exceeds 98% of the low range span. The unit will
return from high range back to low range once both the NO and NO
X
concentrations fall
below 75% of the low range span.
In
AUTO
range mode the
RANGE
test function displayed on the front panel will be replaced
by two separate functions:
•
RANGE1
: The
LOW
range setting for all analog outputs.
•
RANGE2
: The
HIGH
range setting for all analog outputs.
The high/low range status is also reported through the external, digital status bits
(Section 6.9.1). To set individual ranges press the following keystroke sequence.
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL DUAL
AUTO
ENTR EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
AUTO
SNGL
DUAL AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
Toggle the numeral
keys to set the
LOW
and
HIGH
range value.
ENTR accepts the
new setting, EXIT
ignores the new
setting.
NOTE
Low and high ranges have separate slopes and offsets for computing the NO and
NOx concentrations. Hence, the two ranges must be calibrated independently.
6.5.6. Range Units
The M200E can display concentrations in parts per billion (10
9
mols per mol, PPB), parts per
million (10
6
mols per mol, PPM), micrograms per cubic meter (µg/m
3
, UG) or milligrams per
cubic meter (mg/m
3
, MG). Changing units affects all of the display, COM port and iDAS
values for all reporting ranges regardless of the analyzer’s range mode. To change the
concentration units:
52
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...