Model 200E Instruction Manual Instrument
Maintenance
an indication of the lifetime of the chemical in your application. The maintenance
cycle of this item is dependent on ambient moisture, sub-micron particle load and
other factors and may differ from that shown in Table 9-1.
•
Discard the used chemical without touching it. It may contain nitric acid, which is a
corrosive and highly irritating substance. Immediately wash you hands after contact.
•
Using a small powder funnel, fill the cartridge with about 10 g new chemical (TAPI
part # CH43) so that it is level on both legs of the cartridge. Slight vibration is
required to settle the chemical into the cartridge and achieve tightest packing, which
increases performance and lifetime of the filter. Make sure that the level of the
chemical does not protrude farther than the first two threads of the NPT fitting.
NOTE
Use only genuine, pre-conditioned Teledyne-API refill kits for this procedure.
Do not leave this material uncovered for more than a few seconds, as it will
absorb contaminants from ambient air!
•
Ensure that the two fittings still contain a sintered filter at the bottom (pressed in)
and apply a layer of Teflon tape to the NPT threads, then screw the fittings into the
cartridge. Do not over-tighten the fittings, which will damage the soft Teflon threads,
rendering the cartridge useless. The cartridge material will provide a natural seal.
•
Screw the cartridge back onto the chassis, orientation is not important.
•
Evaluate the ferrules on the tubing. If the ferrules are too old, we recommend to
replace them with new ferrules, part # FT 303.
•
Reconnect the tubing.
Do not over-tighten these fittings
. Use only tool # FT267
for half to one more turn beyond hand-tightness. Tools like pliers will destroy the
knurling and may easily over-tighten the fitting and cause it to break.
•
If the service interval for this item has been exceeded, it may also be necessary to
clean the reaction cell as described in Section 9.3.9.
•
Leak check the system using the pressurized approach described in Section 11.5.2.
You can use soap leak detector liquid to evaluate if the cartridge is not tight. If
necessary, tighten the fittings some more but do not overtighten.
•
Restart the analyzer and pump and continue operation.
•
If the chemical used for replacement has not been conditioned properly or was
stored over long periods of time or was exposed to ambient air, you may see a
significant increase of the AutoZero value, slowly diminishing over time. Until
diminished, this will cause large negative offsets on calibration and a changing
calibration over time. It may take one week to clean the material properly, hence,
we recommend to use genuine Teledyne-API refill kits to ensure minimum
downtimes.
•
Re-calibrate the analyzer after one hour (Chapter 7). If AutoZero is high, you may
have to wait day until calibration.
9.3.4. Rebuilding the External Sample Pump
The sample pump head periodically wears out and must be replaced when the RCEL
pressure exceeds 10 in-Hg-A (at sea level, adjust this value accordingly for elevated
044100102 Rev A
147
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...