Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
7.6
Calibrate on NO
2
Permeation Tube
If the M200A is equipped with the IZS option, is possible to span the instrument on the
permeation tube. If this feature is enabled, the instrument internal valves are set so that gas is
routed through the permeation tube, then the moly converter, thus producing NO gas. The
software is programmed to use this known NO gas concentration to span both the NO and NO
x
channels of the instrument. Thus, if CAL-ON-NO2 is enabled, the CALS button now will span
both the NO and NO
x
channels of the instrument.
CAUTION
This feature has some unexpected side effects.
tration
us
tration of NO gas in the NO and NOx channels, and
reports a NO concentration of zero.
disabled so that NO
2
concentrations are reported correctly. See Table 7-8 to enable this feature.
If CAL-ON-NO2 is enabled and you do a SPAN CHECK or SPAN CAL, the NO
2
concen
will be reported as ZERO! This is because all of the NO
2
gas is being routed through the
molybdenum converter during both the NO and NO
x
portions of the measurement cycle. Th
the instrument sees the SAME concen
2
As long as CAL-ON-NO2 is enabled the instrument will never report any NO
2
during span
calibration. This feature should be enabled only when a calibration is desired, otherwise leave it
NOTE
Note that this method of calibration is NOT approved by the USEPA.
Permeation tubes are known to change as temperatures are cycled,
and as they age. Also, this calibration relies on the moly converter
to be working properly. (Converter efficiency can be compensated
by enabling the Converter Efficiency option.) It is recommended that
if CAL-ON-NO2 is used, the instrument calibration be checked
frequently by independent means to assure accurate data.
7-9
Содержание 200A
Страница 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Страница 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Страница 16: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 2 Rear Panel 2 4 ...
Страница 17: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 3 Inlet and Exhaust Venting 2 5 ...
Страница 22: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 4 Front Panel 2 10 ...
Страница 23: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 5 Assembly Layout 2 11 ...
Страница 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Страница 32: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 4 1 Block Diagram 4 2 ...
Страница 38: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 4 2 External Pump Pack 4 8 ...
Страница 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Страница 42: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 5 1 Sample Menu Tree 5 2 ...
Страница 43: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 5 2 Setup Menu Tree 5 3 ...
Страница 91: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 7 1 Calibration Setup 7 3 ...
Страница 109: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 7 2 Diagram of GPT Calibration System 7 21 ...
Страница 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Страница 130: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 1 Replacing the Particulate Filter 8 4 ...
Страница 134: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 2 Molybdenum Converter Assembly 8 8 ...
Страница 140: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 6 Pneumatic Diagram with IZS Option 8 14 ...
Страница 167: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 1 Span Calibration Voltage 9 25 ...
Страница 178: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 2 CPU Board Jumper Settings 9 36 ...
Страница 182: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 3 RS 232 PIN Assignments 9 40 ...
Страница 188: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 4 V F Board Settings 9 46 ...
Страница 193: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 5 Power Supply Module Layout 9 51 ...
Страница 194: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 6 Electrical Block Diagram 9 52 ...
Страница 198: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 7 Ozone Generator Subsystem 9 56 ...
Страница 201: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 8 Flow Pressure Sensor 9 59 ...
Страница 202: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 9 NOx Sensor Module 9 60 ...
Страница 203: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 10 NOx Sensor Module 9 61 ...
Страница 205: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 11 PMT Cooler Subsystem 9 63 ...
Страница 207: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 12 High Voltage Power Supply 9 65 ...
Страница 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Страница 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...