7.8.7 Calibration Frequency
To ensure accurate measurements of the NO, NO
x
, and NO
2
concentrations, calibrate the
analyzer at the time of installation, and re-calibrate it:
1.
No later than three months after the most recent calibration or performance audit which
indicated analyzer calibration to be acceptable.
2.
An interruption of more than a few days in analyzer operation.
3.
Any repairs which might affect its calibration.
4.
Physical relocation of the analyzer.
5.
Any other indication (including excessive zero or span drift) of possible significant
inaccuracy of the analyzer.
Following any of the activities listed above, the zero and span should be checked to determine if
a calibration is necessary. If the analyzer zero and span drifts exceed the calibration limits in
Table 9-1 of Section 2.0.9, Subsection 9.1.3 (Q.A. Handbook), a calibration should be
performed.
7.8.8
Other Quality Assurance Procedures
Precision is determined by a one-point check at least once every two weeks. Accuracy is
determined by a three-point audit once each quarter.
Essential to quality assurance are scheduled checks for verifying the operational status of the
monitoring system. The operator should visit the site at least once each week. Every two weeks a
Level 1 zero and span check must be made on the analyzer. Level 2 zero and span checks should
be conducted at a frequency desired by the user. Definitions of these terms are given in Table 7-
16.
In addition, an independent precision check between 0.08 and 0.10 ppm must be carried out at
least once every two weeks. Table 7-17 summarizes the quality assurance activities for routine
operations. A discussion of each activity appears in the following sections.
To provide for documentation and accountability of activities, a checklist should be compiled
and then filled out by the field operator as each activity is completed.
For information on shelter and sample inlet system, an in-depth study is in Field Operations
Guide for Automatic Air Monitoring Equipment, Publication No. APTD-0736, PB 202-249 and
PB 204-650, U.S. Environmental Protection Agency, Office of Air Programs, October 1972.
7-27
Содержание 200A
Страница 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Страница 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Страница 16: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 2 Rear Panel 2 4 ...
Страница 17: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 3 Inlet and Exhaust Venting 2 5 ...
Страница 22: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 4 Front Panel 2 10 ...
Страница 23: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 5 Assembly Layout 2 11 ...
Страница 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Страница 32: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 4 1 Block Diagram 4 2 ...
Страница 38: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 4 2 External Pump Pack 4 8 ...
Страница 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Страница 42: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 5 1 Sample Menu Tree 5 2 ...
Страница 43: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 5 2 Setup Menu Tree 5 3 ...
Страница 91: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 7 1 Calibration Setup 7 3 ...
Страница 109: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 7 2 Diagram of GPT Calibration System 7 21 ...
Страница 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Страница 130: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 1 Replacing the Particulate Filter 8 4 ...
Страница 134: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 2 Molybdenum Converter Assembly 8 8 ...
Страница 140: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 6 Pneumatic Diagram with IZS Option 8 14 ...
Страница 167: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 1 Span Calibration Voltage 9 25 ...
Страница 178: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 2 CPU Board Jumper Settings 9 36 ...
Страница 182: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 3 RS 232 PIN Assignments 9 40 ...
Страница 188: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 4 V F Board Settings 9 46 ...
Страница 193: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 5 Power Supply Module Layout 9 51 ...
Страница 194: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 6 Electrical Block Diagram 9 52 ...
Страница 198: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 7 Ozone Generator Subsystem 9 56 ...
Страница 201: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 8 Flow Pressure Sensor 9 59 ...
Страница 202: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 9 NOx Sensor Module 9 60 ...
Страница 203: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 10 NOx Sensor Module 9 61 ...
Страница 205: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 11 PMT Cooler Subsystem 9 63 ...
Страница 207: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 12 High Voltage Power Supply 9 65 ...
Страница 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Страница 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...