2074
User Manual
7-4
Recommended
Test Equipment
Recommended equipment for adjustments is listed in Table 7-1.
Instruments other than those listed may be used only if their
specifications equal or exceed the required minimal characteristics.
Also listed below are accessories required for calibration.
Table
7-1, Recommended calibration for Adjustments
Equipment
Model No.
Manufacturer
Oscilloscope (with jitter package)
LC684
LeCroy
Digital Multimeter
2000
Keithley
Frequency Counter (Rubidium reference) 6020R
Tabor Electronics
Function Generator (with manual trigger)
8020
Tabor Electronics
Accessories
BNC to BNC cables
50
Ω
Feedthrough termination
Dual banana to BNC adapter
Adjustment
Procedures
Use the following procedures to calibrate the Model 2074. The
following paragraphs show how to set up the instrument for calibration
and what the acceptable calibration limits are.
Calibration is done with the covers closed and the 2074 connected
through an interface to a host computer. Any interface can be used
from the following: USB, LAN, or GPIB. Calibration requires that
ArbConnection utility be installed and interfaced to the instrument.
Calibration is performed from the Calibration Panel in ArbConnection.
To invoke this panel, one requires a password that is available to
service centers only. Contact your nearest Tabor service center for
information and permit to obtain your calibration password. Use the
following procedure to calibrate the generator:
1. Invoke
ArbConnection
2. Click on the Calibration tab on the Panels bar
3. Expect to be prompted with the following dialog box
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...