User Manual
6-20
(n)QAM
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
1 ms
Sampling Rate: 50 MS/s at least.
Trace View:
X-Y
Amplitude:
1 V/div
2. Connect 2074 Channel 1 output to the oscilloscope input,
channel 1
3. Connect the 2074 Channel 2 output to the oscilloscope input,
channel 2
4. Configure model 2074 controls on channels 1 and 2 as follows:
Waveform:
Modulated
Modulation:
nQAM
QAM Type:
16QAM
CW Control:
Off
CH1 Output:
On
CH2 Output:
On
Test Procedure:
1. Load Demo table to the 16QAM Data Table
2. Set oscilloscope and verify 16QAM constellation trace on the
oscilloscope
Test Results
Pass
Fail
6. Modify 2074 QAM Type to 64QAM
7. Load Demo table to the 64QAM Data Table
8. Verify 64QAM constellation trace on the oscilloscope
Test Results
Pass
Fail
SYNC Output
operation
This tests the operation of the SYNC output. There are four separate
synchronization outputs, which are located on the rear panel. There
are two parameters being tested, the sync qualifier and the sync
position. The sync output has TTL level amplitude.
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...