5-62
AUXiliary:PULse:POLarity{NORMal||COMPlemented|INVerted (?)
Description
This command will program the polarity of the pulse in reference to the base line level. The polarity options
are: Normal, where the pulse is generated exactly as programmed; Inverted, where the pulse is inverted
about the 0 level base line; and Complemented, where the pulse is inverted about its mid amplitude level.
Parameters
Name
Type
Default
Description
NORMal
Discrete
NORM
Programs normal pulse output
COMPlemeted
Discrete
Programs complemented pulse output
INVerted
Discrete
Programs an inverted pulse output
Response
The 2074 will return NORM, COMP or INV depending on the present polarity setting
AUXiliary:PULse{OFF|ON|0|1}(?)
Description
Use this command to disable a specific channel from calculating pulse parameters. This is specifically
useful for accelerating pulse computation for channels that are needed for pulse generation.
Parameters
Range
Type
Default
Description
0-1
Discrete
0
Toggles pulse computation for a specific channel on and
off
Response
The 2074 will return 0, or 1 depending on the present state setting.
AUXiliary:PULse:TRANsition<rise>(?)
Description
This command will program the interval it will take the pulse to transition from its low to high level settings.
The parameter is programmed in units of seconds.
Parameters
Name
Range
Type
Default
Description
<rise>
0 to 1e3
Numeric
1e-3
Will set the rise time parameter. Note that the sum of
all parameters, including the rise time must not
exceed the programmed pulse period and therefore,
it is recommended that the pulse period be
programmed before all other pulse parameters.
Response
The 2074 will return the present rise time value in units of seconds.
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...