going transitions will trigger the generator when the POS option is selected. Negative transitions will trigger
the generator when the NEG option is selected. In Gated mode, two transitions in the same direction are
required to gate on and off the output. The trigger slope command will affect the generator only after it has
been programmed to operate in interrupted run mode. Modify the 2074 to interrupted run mode using the
init:cont off command.
Parameters
Name
Type
Default
Description
POSitive
Discrete
POS
Selects the positive going edge.
NEGative
Discrete
Selects the negative going edge.
Response
The 2074 will return POS, or NEG depending on the selected trigger slope setting.
RETRigger{OFF|ON|0|1}(?)
Description
This command will toggle the re-trigger mode on and off. This command will affect the 2074 only after it will
be set to INIT:CONT OFF.
Parameters
Name
Type
Default
Description
OFF
Discrete
OFF
Turns the re-trigger mode off.
ON
Discrete
Enables the re-trigger mode.
Response
The 2074 will return OFF, or ON depending on the selected option.
RETRigger:Time<time>(?)
Description
This parameter specifies the amount of time that will elapse between the end of the delivery of the waveform
cycle and the beginning of the next waveform cycle. Re-trigger can be initiated from any of the selected
advance options. The re-trigger command will affect the generator only after it has been programmed to
operate in interrupted run mode. Modify the 2074 to interrupted run mode using the init:cont off command.
Parameters
Name
Range
Type
Default
Description
<time>
200e-9 to 20
Numeric
200e-9
Programs the re-trigger period.
Response
The 2074 will return the present re-trigger period value.
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...