5-20
FREQuency:RASTer{<sclk>|MINimum|MAXimum}(?)
Description
This command modifies the sample clock frequency of the arbitrary waveform in units of samples per
second (S/s). It has no affect on standard waveforms.
Parameters
Name
Range
Type
Default
Description
<sclk>
1.5 to
200e6
Numeric
1e7
Will set the sample clock frequency of the arbitrary
and sequenced waveform in units of S/s. Although
the display resolution for the frequency setting is 9
digits only, the frequency command can be used with
resolutions up to 14 digits. The accuracy of the
instrument however, can only be tested to this
accuracy using an external reference that provides
the necessary accuracy and stability
<MINimum>
Discrete
Will set the sample clock frequency to the lowest
possible frequency (1.5).
<MAXimum>
Discrete
Will set the frequency of the standard waveform to
the highest possible frequency (200e6).
Response
The 2074 will return the present sample clock frequency value. The returned value will be in standard
scientific format (for example: 100MHz would be returned as 100e6
– positive numbers are unsigned).
FREQuency:RASTer:SOURce{EXTernal|INTernal}(?)
Description
This command selects the source of the sample clock generator. This command affects both the standard
and the arbitrary waveforms.
Parameters
Name
Type
Default
Description
INTernal
Discrete
INT
Selects an internal source.
EXTernal
Discrete
Activates the external sample clock reference input.
An external reference must be connected to the
2074, in the range of the internal source, for it to
continue normal operation. Observe the input level
and limitations before connecting an external signal.
Response
The 2074 will return EXT if an external source is selected, or INT if the internal source is selected.
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...