Chapter 8: Operating Modes
Oximeter Clinician’s Operation Manual
8-1
Chapter 8: Operating Modes
About the Monitor’s Operating Modes
The monitor has three operating modes: clinician mode (previously
described), home-use mode and sleep study mode.
•
The clinician mode is intended for health-care professionals
trained in monitoring respiratory and cardiovascular activity.
•
The home-use mode is intended for caregivers trained in oxime-
ter use by a doctor or other health-care professional.
•
The sleep study mode is intended for health-care professionals
trained in monitoring respiratory and cardiovascular activity.
While the monitor is in the clinician mode, all monitor functions operate
as described previously in this manual.
NOTE! Accessing Home Use Mode and Sleep Study Mode may
only be done from the Clinician Mode. It is not possible to
access Home Use Mode from Sleep Study Mode or vice versa.
Home Use Mode
While the monitor is in home-use mode:
•
Alarm limits can be viewed but cannot be changed.
•
The alarm and alert tone is set to loud and cannot be changed.
•
Alarm and alert tones can be silenced for two minutes but can-
not be silenced indefinitely.
•
SpO
2
averaging is set to 8 beats; pulse rate averaging is set to 8
seconds.
•
Patient numbers and trend view are disabled.
•
Trend data is collected for one patient.
•
Trend data from the patient can be printed. The printout is the
same as during clinician mode, except “Home Mode” replaces
“Patient Number” on the printout.
•
All other functions of the monitor work as in clinician mode.
Содержание 1850
Страница 2: ......
Страница 4: ......
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 22: ...This page is intentionally left blank ...
Страница 30: ...This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank ...
Страница 70: ...This page is intentionally left blank ...
Страница 81: ......