Pulse Rate
Range
30-254 BPM
Accuracy
±2% at 30-254 BPM
Alarm Limits
High 250-5 BPM and off in 5 BPM
steps.
Low 5-250 BPM and off in 5 BPM
steps.
Averaging
8 or 16 second average.
Alarm Indicators
Two-tone audible alarm with user-adjustable vol-
ume and two-minute or indefinite alarm silence.
Corresponding numeric display flashes.
Sensor Alert Indicator
Single-tone audible alarm with same volume and
silence as alarm tone.
Printer Output
Data saved every thirty (30) seconds can be
printed (trend).
Battery
Type
Internal rechargeable NiCad or
NiMH; not user replaceable.
Charge Time
Nicad: Fully charges in about 4
hours.
NiMH: Fully charges in about 6
hours.
Use Time
Approximate continuous use:
12 hours (Nicad)
24 hours (NiMH).
Indicators
BATT indicator lights when
about 30 minutes of battery use
remains.
AC Charger
Wall Mount
Style
Input of 105-125 V~, 60Hz
Input of 100V~, 50Hz (Optional)
Table Top Style Input of 230 V~, 50/60Hz
Dimensions
Width
3.3 inches (84 mm)
Height
7.25 inches (184 mm)
Depth
1.85 inches (47 mm)
Weight
19 ounces (539 grams) including
battery
Environmental Specifica-
tions
Temperature
Operating:-20-55° C (-4-131° F)
Storage: -40-75° C (-40-167° F)
Relative
Humidity
Operating: 15-95%, non-con-
densing
Storage: 10-90%, non-condens-
ing
Chapter 13: Specifications
13-2
Oximeter Clinician’s Operation Manual
Содержание 1850
Страница 2: ......
Страница 4: ......
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 22: ...This page is intentionally left blank ...
Страница 30: ...This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank ...
Страница 70: ...This page is intentionally left blank ...
Страница 81: ......