Figure 3-2: Front Panel
6.
Sensor
Light
7.
BATT
Light
5.
Pulse
Strength
Bar Graph
Chapter 3: Controls
3-2
Oximeter Clinician’s Operation Manual
5. Pulse Strength Bar Graph
The pulse strength bar graph “sweeps” with the patient’s pulse beat. The
height of the bar graph tells the strength of the patient’s pulse.
6. Sensor Light
SENSOR flashes on and off when the sensor is not connected to the
monitor, the sensor is not attached to the patient, or the sensor is not
properly attached to the patient.
2
WARNING! While SENSOR is flashing, the monitor cannot measure the
patient’s SpO
2
or pulse rate. You must immediately check the patient’s
condition. After you have checked the patient’s condition, you must
correct the SENSOR alert.
7. BATT Light
BATT flashes on and off when about 30 minutes of battery use remains.
The monitor will work until the battery becomes very weak. When the
battery becomes very weak, the monitor will turn itself off.
2
WARNING! When BATT flashes, you must immediately charge
the monitor’s battery. Otherwise, the monitor turns itself off
about 30 minutes after BATT begins to flash.