Chapter 4: Operating Instructions
Oximeter Clinician’s Operation Manual
4-15
BATT Attention
Figure 4-10: BATT Attention Example
BATT flashes on and off dur-
ing a BATT Attention
During the BATT attention:
•
a short burst of beeps sounds every 30 seconds.
•
the BATT light flashes on and off.
2
WARNING! When BATT flashes, you must immediately charge
the monitor’s battery. Otherwise, the monitor turns itself off
about 30 minutes after BATT begins to flash. See
BATT atten-
tion
in the
Troubleshooting
section for help.
Содержание 1850
Страница 2: ......
Страница 4: ......
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 22: ...This page is intentionally left blank ...
Страница 30: ...This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank ...
Страница 70: ...This page is intentionally left blank ...
Страница 81: ......