Chapter 7: Printer
Oximeter Clinician’s Operation Manual
7-5
Trend Printouts
Collecting Trend Data
Whenever the monitor is on, it stores one SpO
2
and one pulse rate read-
ing every thirty (30) seconds. The stored readings are called
trend
data.
The monitor remembers trend data for up to 99 patients and 24 hours of
run-time. The trend data can then be printed at any time on the optional
printer.
Trend data is saved for each patient number. When you turn on the mon-
itor, the patient number is automatically incremented and displayed dur-
ing the power-up sequence if valid trend data was collected from the
previous patient. If no valid trend data was collected from the previous
patient, the patient number is displayed only and is not incremented.
The monitor remembers all the trend data and all the patient numbers
for up to 99 patients and 24 hours of run-time.
Manually Incrementing the Patient Number
1. The SpO
2
sensor must be connected to the monitor. If necessary,
connect the SpO
2
sensor to the monitor.
2. Press the I.D./CLEAR key to increment the patient number. The new
patient number is momentarily displayed and trend data for the new
patient is automatically saved.
Clearing Trend Data
1. The SpO
2
sensor must be connected to the monitor. If necessary,
connect the SpO
2
sensor to the monitor.
2. Press and hold the I.D./CLEAR key for about six seconds. While you
are holding the I.D./CLEAR key, the message
[{r
flashes on the dis-
play to tell you the trend data for all patients is about to be cleared.
When the trend data is cleared, the display shows
P1
.
Содержание 1850
Страница 2: ......
Страница 4: ......
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 22: ...This page is intentionally left blank ...
Страница 30: ...This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank ...
Страница 70: ...This page is intentionally left blank ...
Страница 81: ......