Chapter 8: Operating Modes
Oximeter Clinician’s Operation Manual
8-9
b. Turn on the monitor while observing the display;
SLP
should be shown briefly when the monitor is turned on. If
not, repeat step 1 above.
Turning Off Sleep Study Mode
1. Turn off the sleep study mode (put the monitor into the clinician
mode) as follows:
a.
Turn off the monitor.
b. Press and hold the Alarm Silence key, then press the ON key.
c.
While holding the Alarm Silence key,
SLP
flashes in the
pulse rate display.
d. When
SLP
stops flashing, and patient number
pn
(n =
patient number) lights steady, (in about 6 seconds), release
the key. The monitor is now in clinician mode.
2. Verify the monitor is in the clinician mode:
a.
Turn off the monitor.
b. Turn on the monitor while observing the display; the soft-
ware revision then the patient number should be displayed
during power-up. If not, repeat step 1 above.