NOTE! “ – – – ” in the display means the limit is set to off.
NOTE! Alarm limits are non-overlapping. You cannot set the
high alarm equal to or lower than the low alarm and you can-
not set the low alarm equal to or higher than the high alarm.
NOTE! While setting alarm limits, if no keys are pressed for
twenty seconds, the alarm limit setting mode is exited and the
SpO
2
and pulse rate measurements are shown.
NOTE! Alarms are not active while setting alarm limits; how-
ever, alarms are active as soon as you exit the alarm limit set-
ting mode.
NOTE! The low SpO
2
alarm limit minimum test value is 80. If an
operator changes the low SpO
2
alarm limit to a value less than
80, and a power down - power up sequence takes place, a mini-
mum value of 85 takes the place of the operator entered value.
Chapter 5: Changing the Monitor’s Settings
Oximeter Clinician’s Operation Manual
5-3