NOTE! Operation of this device may be adversely affected in
the presence of strong electromagnetic sources, such as elec-
trosurgery equipment.
NOTE! Operation of this device may be adversely affected in
the presence of computed tomography (CT) equipment.
NOTE! Use only SpO
2
sensors supplied with, or specifically
intended for use with, this device
NOTE! SpO
2
measurements may be adversely affected in the
presence of high ambient light. Shield the sensor area (with a
surgical towel, for example) if necessary.
NOTE! Dyes introduced into the bloodstream, such as methyl-
ene blue, indocyanine green, indigo carmine, patent blue V
(PBV), and fluorescein may adversely affect the accuracy of the
SpO
2
reading.
NOTE! Any condition that restricts blood flow, such as use of a
blood pressure cuff or extremes in systemic vascular resis-
tance, may cause an inability to determine accurate SpO
2
and
pulse rate readings.
NOTE! Remove fingernail polish or false fingernails before
applying SpO
2
sensors. Fingernail polish or false fingernails
may cause inaccurate SpO
2
readings.
NOTE! Batteries are not user replaceable.
Service Personnel: Follow local governing ordinances and
recycling instructions regarding disposal or recycling of device
components, including batteries.
NOTE! Optical cross-talk can occur when two or more sensors
are placed in close proximity. It can be eliminated by covering
each site with an opaque material.
Chapter 1: Introduction
Oximeter Clinician’s Operation Manual
1-9
Содержание 1850
Страница 2: ......
Страница 4: ......
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 22: ...This page is intentionally left blank ...
Страница 30: ...This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank ...
Страница 70: ...This page is intentionally left blank ...
Страница 81: ......