Chapter 8: Operating Modes
8-2
Oximeter Clinician’s Operation Manual
Setting Up the Monitor for Home-Use
1. Set the high and low alarm limits for SpO
2
and pulse rate to the val-
ues prescribed by the doctor:
•
High SpO
2
alarm limit.
•
Low SpO
2
alarm limit.
•
High pulse rate alarm limit.
•
Low pulse rate alarm limit.
2. Clear trend data if necessary. (Press and hold I.D./CLEAR until P1 is
displayed.)
3. Put the monitor into the home-use mode as follows (Make sure that
the monitor is in Clinician Mode prior to executing the following
commands!):
a.
Turn off the monitor.
b. Press and hold the ALARM VOL key, then press the ON key.
c.
While holding the ALARM VOL key,
k
flashes in the pulse
rate display.
d. When
k
stops flashing and lights steady (in about 6 sec-
onds), release the ALARM VOL key. The monitor is now in
home-use mode.
4. Verify the monitor is in the home-use mode:
a.
Turn off the monitor.
b. Turn on the monitor while observing the display;
k
should
be shown briefly when the monitor is turned on. If not,
repeat step 3 above.
5. Verify the high and low alarm limits for SpO
2
and pulse rate are set to
the values prescribed by the doctor. (Press ALARM SEL to show each
of the alarm limits one-at-a-time.)
6. Turn off the monitor.
Содержание 1850
Страница 2: ......
Страница 4: ......
Страница 8: ...This page is intentionally left blank ...
Страница 18: ...This page is intentionally left blank ...
Страница 22: ...This page is intentionally left blank ...
Страница 30: ...This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank ...
Страница 70: ...This page is intentionally left blank ...
Страница 81: ......