Calibrating 20 mA Output
1.
Connect the mA receiving device to the device.
2.
Put the device into Navigation View.
3.
Set Milliamp Output (2.5.9.) (Page 182) to 20.0 for the selected mA output.
4.
View the mA level on the receiving device.
5.
If there is a discrepancy, do the following steps:
–
Attach an ammeter to the device's mA output.
–
Enter the exact value displayed on the ammeter into 20 mA Output Trim (2.5.12.)
(Page 183)
–
The ammeter should then read exactly 20.00 mA.
The unit is now calibrated for 20 mA for the receiving device.
7.7
Volume
Volume is used in two situations:
1.
To calculate and display Volume instead of Level.
2.
To calculate pumped volume to accomplish the following:
●
Totalize the volume of material that is pumped out of the wet well
●
Set an alarm on pump efficiency
7.7.1
Readings
When using volume, volume values are presented in the same units specified in Maximum
Volume (2.7.3.) (Page 188). The default is 100, which gives a reading in percent of total. Use
whatever unit is applicable. If the value is too large for the six-digit LCD, use a larger unit.
Содержание HydroRanger 200 HMI
Страница 2: ......
Страница 20: ...Introduction 1 5 Notes on warranty HydroRanger 200 HMI 18 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 24: ...Safety notes HydroRanger 200 HMI 22 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 28: ...Description 3 5 Modbus communication HydroRanger 200 HMI 26 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 159: ...Parameter reference 8 2 Parameter indexing HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 157 ...
Страница 276: ...Parameter reference 8 10 Language 6 HydroRanger 200 HMI 274 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 322: ...Pump control reference B 13 Other pump controls HydroRanger 200 HMI 320 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 352: ...Communications C 41 Single parameter access SPA HydroRanger 200 HMI 350 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 354: ...Updating software HydroRanger 200 HMI 352 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 359: ...HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 357 Conduit entry for Class I Div 2 applications F ...
Страница 360: ...Conduit entry for Class I Div 2 applications HydroRanger 200 HMI 358 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 361: ...Conduit entry for Class I Div 2 applications HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 359 ...
Страница 362: ......
Страница 372: ...Programming chart G 1 Programming chart HydroRanger 200 HMI 370 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 390: ...LCD menu structure H 1 LCD Menu Structure HydroRanger 200 HMI 388 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 403: ......