8.6.5.10
Fail-safe mode (2.5.10.)
Use for fail-safe operation, independent of the Material Level (2.4.5.) (Page 178).
Index
mA output
Options
*OFF
mA output responds to Material Level (2.4.5.) (Page 178)
HI
Produce the Span mA output immediately.
LO
Produce the Empty mA output immediately.
Hold
Last known value is held until normal operation resumes.
Value
mA output goes to user-selected value defined in mA Fail-
safe Value (2.4.9.) (Page 180).
Related
Current output function (2.5.3.) (Page 181)
8.6.5.11
4 mA output trim (2.5.11.)
Calibrates the 4 mA output
Index
mA Output
Values
Preset: 4.00
Related
20 mA output trim (2.5.12.) (Page 183)
Adjust this value so the device indicates 4.00 mA when 4 mA Output Trim (2.5.11.) is
accessed.
Trimming the 4 mA value:
1.
Go to Milliamp Output (2.5.9.) (Page 182) and enter 4.00 mA.
2.
Read the value indicated on the current meter or other connected device, and enter it in
mA output trim (2.5.11.)
The new trimmed value will be shown in the current meter or other connected device.
8.6.5.12
20 mA output trim (2.5.12.)
Calibrates the 20 mA output
Index
mA Output
Values
Preset: 20.00
Related
4 mA Output Trim (2.5.11.) (Page 183)
Adjust this value so the device indicates 20.00 mA when 20 mA output trim (2.5.12.) is
accessed
Trimming the 20 mA value:
1.
Go to Milliamp Output (2.5.9.) (Page 182) and enter 20.00 mA.
2.
Read the value indicated on the current meter or other connected device, and enter it in
mA output trim (2.5.12.)
The new trimmed value will be shown in the current meter or other connected device.
Содержание HydroRanger 200 HMI
Страница 2: ......
Страница 20: ...Introduction 1 5 Notes on warranty HydroRanger 200 HMI 18 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 24: ...Safety notes HydroRanger 200 HMI 22 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 28: ...Description 3 5 Modbus communication HydroRanger 200 HMI 26 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 159: ...Parameter reference 8 2 Parameter indexing HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 157 ...
Страница 276: ...Parameter reference 8 10 Language 6 HydroRanger 200 HMI 274 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 322: ...Pump control reference B 13 Other pump controls HydroRanger 200 HMI 320 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 352: ...Communications C 41 Single parameter access SPA HydroRanger 200 HMI 350 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 354: ...Updating software HydroRanger 200 HMI 352 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 359: ...HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 357 Conduit entry for Class I Div 2 applications F ...
Страница 360: ...Conduit entry for Class I Div 2 applications HydroRanger 200 HMI 358 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 361: ...Conduit entry for Class I Div 2 applications HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 359 ...
Страница 362: ......
Страница 372: ...Programming chart G 1 Programming chart HydroRanger 200 HMI 370 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 390: ...LCD menu structure H 1 LCD Menu Structure HydroRanger 200 HMI 388 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 403: ......