The Echo Lock Window is a distance window [Units (2.1.1.) (Page 162) ] centered on the
echo and used to derive the Reading. When a new measurement is in the window, it is re-
centered and the new reading calculated. Otherwise, the new measurement is verified by
Echo Lock (2.11.5.4.) (Page 235) before the reading is updated.
When 0 is entered, the window is automatically calculated after each measurement. For
slower Response Rate (2.3.4.) (Page 174) values, the window is narrow; for faster Response
Rate (2.3.4.) (Page 174) values, the window becomes wider.
Fuzz filter (2.11.5.6.)
Stabilizes the reported level, due to level fluctuations (such as a rippling or splashing liquid
surface) within the Echo Lock Window (2.11.5.5.) (Page 235).
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0 ... 100 (0 = OFF)
Altered by
Response rate (2.3.4.) (Page 174)
Related
•
•
Echo lock window (2.11.5.5.) (Page 235)
This value [in % of Span (2.2.2.) (Page 169)] is automatically altered when Response Rate
(2.3.4.) (Page 174) is changed. The higher the value entered, the greater the fluctuation
stabilized.
8.6.12
Display (2.12.)
8.6.12.1
Local display backlight (2.12.1)
Controls the LCD backlighting.
Index
Global
Options
OFF
*ON
Timed
The backlight can be forced ON or OFF, or be controlled by a programmer, in which case it
will turn OFF five minutes after the last key is pressed.
8.6.12.2
LCD contrast (2.12.2.)
Adjusts the luminance of the bright and dark colors of the display.
Options
Range: 0 (Low contrast) to 20 (High contrast)
Preset: 10
Содержание HydroRanger 200 HMI
Страница 2: ......
Страница 20: ...Introduction 1 5 Notes on warranty HydroRanger 200 HMI 18 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 24: ...Safety notes HydroRanger 200 HMI 22 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 28: ...Description 3 5 Modbus communication HydroRanger 200 HMI 26 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 159: ...Parameter reference 8 2 Parameter indexing HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 157 ...
Страница 276: ...Parameter reference 8 10 Language 6 HydroRanger 200 HMI 274 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 322: ...Pump control reference B 13 Other pump controls HydroRanger 200 HMI 320 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 352: ...Communications C 41 Single parameter access SPA HydroRanger 200 HMI 350 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 354: ...Updating software HydroRanger 200 HMI 352 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 359: ...HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 357 Conduit entry for Class I Div 2 applications F ...
Страница 360: ...Conduit entry for Class I Div 2 applications HydroRanger 200 HMI 358 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 361: ...Conduit entry for Class I Div 2 applications HydroRanger 200 HMI Operating Instructions 06 2018 A5E36281317 AC 359 ...
Страница 362: ......
Страница 372: ...Programming chart G 1 Programming chart HydroRanger 200 HMI 370 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 390: ...LCD menu structure H 1 LCD Menu Structure HydroRanger 200 HMI 388 Operating Instructions 06 2018 A5E36281317 AC ...
Страница 403: ......