Functions
Start pos
Specifies by how much the storage start point should be shifted from the
point where the start condition is satisfied.
The unit is number of samplings.
• Positive value: Delays by the specified sampling amount and then starts
storage. (Trigger delay)
• 0: Starts storage the instant the condition is satisfied.
• Negative value: Backs up by the specified sampling amount and then
starts storage. (Pre-trigger)
Start target
When "Start cond" is "Measure", "Alarm", "UP limit", or "LOW limit", this
setting selects which area should be the reference of the start condition.
Select from Area1 to 4, or Calc1 or 2.
Threshold
When "Start cond" is "UP limit" or "LOW limit", this setting specifies the
threshold to be used as the start condition reference.
7.3.3
Intermittent
Specifies the sampled measurement results and profile storage interval.
0: Stores all acquired data.
1: Repeatedly stores and discards data with each measurement.
The thinning setting affects the storage function only.
Analog output, judgment output, etc. are updated for each sampling.
7.3.4
Repeat
Specifies behavior when the specified number of data items is stored.
ON: Returns to the beginning and overwrites the currently stored data.
OFF: Storage is stopped when the preset number of data items is reached.
The storage target automatically changes to OFF.
Pause: When the start condition is fulfilled, storage is proceeded automat-
ically up to the preset number of data items and stops. After that, when
the start condition is fulfilled again, it starts storage again from the point
it stopped previously. When it reached at upper limit of the storage data, it
returns to the beginning and overwrites the currently stored data same as
when this setting is “ON”.