Serial Communication
126
© SICK AG • Subject to change without notice. • 8017035/ZMO9/2017-06-08
Function
Set or
acquire
Send
or
reply
Length Command
(hex)
Data string
Parameter
Average
Set
Send
2
8014
000a000b
a: Area
a
Setting value
a
Setting value
0
Area1
2
Area3
1
Area2
3
Area4
bbb: Number of times to perform the moving
average
1 to 1023 [number of times]
Reply
0
—
Acquire Send
1
8015
000a
Reply
1
000b
Hold
Set
Send
2
8016
000a000b
a: Area/calculation
a
Setting value
a
Setting value
0
Area1
3
Area4
1
Area2
4
Calc1
2
Area3
5
Calc2
b: Hold operation
b
Setting value
b
Setting value
0
None
2
Peak
1
Sample
3
Bottom
Reply
0
—
Acquire Send
1
8017
000a
Reply
1
000b
Calc target1
Set
Send
2
8002
000a000b
a: Calculation
a
Setting value
a
Setting value
4
Calc1
5
Calc2
b: Measurement direction
b
Setting value
b
Setting value
0
Area1
2
Area3
1
Area2
3
Area4
Reply
0
—
Acquire Send
1
8003
000a
Reply
1
000b
Operator
Set
Send
2
8006
000a000b
a: Calculation
a
Setting value
a
Setting value
4
Calc1
5
Calc2
b: Operator
b
Setting value
b
Setting value
0
+
1
—
Reply
0
—
Acquire Send
1
8007
000a
Reply
1
000b