Functions
74
© SICK AG • Subject to change without notice. • 8017035/ZMO9/2017-06-08
7.2
Input/Trigger Settings
7.2.1
IN1/IN2/IN3/IN4
Set the operation of the external input terminals. The available selections
vary depending on the terminal. IN1, IN2, and IN3 are used as bits 1 to 3
of a binary number that is used to switch the bank.
Use the sensor menu to switch the functions of terminals that are not used
in bank switching.
on page 90)
IN1
• BANK1:
IN1 will be used as the first bit for switching banks.
• Reset:
IN1 will be used as the reset input terminal.
IN2
• BANK2:
IN2 will be used as the second bit for switching banks.
• Hold:
IN2 will be used as the hold input terminal. The set
hold operation will be performed for the areas and
calculations whose hold settings have been set
to “Sample,” “Peak,” or “Bottom” in the Graph/Calc
setting (“Calculation settings” in PRO2-Navigator).
IN3
• BANK3:
IN3 will be used as the third bit for switching banks.
• Trigger:
IN3 will be used as the trigger input terminal.
IN4
• Offset:
IN4 will be used as offset input.
• LaserOFF:
IN4 will be used as the laser emission stop input.
Input is used to stop laser emission and measurement.
7.2.2
Reset/Inner hold/Inner trig
Reset
The reset operation is executed from the sensor or from PRO2-Navigator.
When you apply the reset input, the following values will be cleared.
• The current measured value
• The measured values up to the current point in time
(when the moving average has been set)
• The held value
• Storage data
Also, when a reset input is applied, the output will enter trigger input stand-
by state until the measured value is obtained.
Inner hold
Hold is turned on or off from the sensor or from PRO2-Navigator.
This is only valid when IN2 is set to “BANK2.”
Inner trig
The trigger is turned on or off from the sensor or from PRO2-Navigator.
This is only valid when IN3 is set to “BANK3.”