Portable X-ray Units
Service Manual
SM-1076R2
67
ERROR CODE :
E45
DESCRIPTION
:
I2C Bus error while trying to access the multiplexer.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
It is not possible to communicate with the multiplexer MAX4586 (U12) (Control Driver
Board
A3189--04
).
ACTIONS
1.
Check on the Control Portable Board (
A3175--03
):
G
I2C Bus pull--up resistors (R66 and R67).
G
Review all the hardware connected to the I2C Bus: P89LPC922 (U56),
M41T56 (U54).
2.
Check on the Control Driver Board (
A3189--04
) the digital potentiometer AD5245
(U10) and the multiplexer MAX4586 (U12).
3.
Review the related components to all of them.
4.
If the error persists, replace the defective hardware.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...