Portable X-ray Units
Service Manual
SM-1076R2
24
5.
Make an exposure and measure mA signal value with the Oscilloscope.
Verify that mA signal is flat and noiseless, make sure that the signal is
balanced before it arrives to 20 ms.
The white line represents the mA Signal
a
b
The signal in both cases (a and b) is balanced before it arrives to 20 ms
6.
If the signal is not balanced before it arrives to 20 ms, change the “
Ifil
”
(
Filament Current
) value with mA and s and repeat the exposure until the
signal is flat before 20ms as shown in the previous picture.
7.
Enter mA read value in the mAs Display pressing
“mAs Increase”
or
“mAs
Decrease”
push-buttons.
8.
Save the value by pressing
“Reset”
push-button. The mAs Display shows
mA Gain value during two seconds (this value represents the Gain
correction automatically applied, 200 is the reference value).
9.
Press
“Collimator Light”
to exit menu P02.
10.
To exit Calibration Mode, press and hold “
Collimator Light”
push-button
and then press and release
“mAs Decrease”
(two beeps sound). Then
release
“Collimator Light”
(kVp and mAs parameters are shown).
P03 and up are not used menus.
Note
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...