Portable X-ray Units
Service Manual
SM-1076R2
90
5.3.14
CONTROL DRIVER BOARD
1.
Access inside the Unit by removing Upper and Lower Covers and
unscrewing the Connectors Panel (
refer to Section 5.3.1
).
2.
Remove the Filament HF Board (
A9500--01
) as described in
Section
5.3.13
.
3.
Disconnect from the Control Driver Board (
A3189--04
) the cables
connected to J1, J2, J3, J4, J5, GND. Then remove the four (4) nuts
attaching the cables connected to TA, TB, TC, TD bolts and take out the
cables from the bolts.
Take note of the correspondence of each cable connected to
each bolt (TA, TB, TC, TD). It is recommended to mark each
cable to avoid any mistakes when replacing the Board.
GND
J1
J5
J3
J2
J4
TD
TC
TB
TA
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...