Portable X-ray Units
Service Manual
SM-1076R2
59
ERROR CODE :
E30
DESCRIPTION
:
Wrong date/time at the RTC (Real Time Clock). Wrong date stored in the RTC (Real
Time Clock) and/or the Time Stamp. “Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The RTC has not been configured, the RTC battery is discharged or missed or the RTC
is failing.
ACTIONS
1.
Check the RTC configuration (
refer to Section 2.2.7
).
2.
If the error persists, check the battery BAT1 (model CR 1216, 3V) of the Portable
Control Board (
A3175--03
) and replace if needed.
3.
If the error persists, replace the Portable Control Board (
A3175--03
) and initialize
the new EEPROM (
refer to Section 5.1.1
).
ERROR CODE :
E31
DESCRIPTION
:
Wrong Time Stamp. The Time Stamp checksum is wrong. “Recoverable” with the “
Reset
”
push-button.
POSIBLE CAUSES
The RTC has not been configured, the RTC battery is discharged or missed or the RTC
is failing.
ACTIONS
1.
Check the RTC configuration (
refer to Section 2.2.7
).
2.
If the error persists, check the battery BAT1 (model CR 1216, 3V) of the Portable
Control Board (
A3175--03
) and replace if needed.
3.
If the error persists, replace the Portable Control Board (
A3175--03
) and initialize
the new EEPROM (
refer to Section 5.1.1
).
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...