Portable X-ray Units
Service Manual
SM-1076R2
7
Table 2-1 (cont.)
Configuration Menus
Value on kVp Display
DESCRIPTION
NOTES
TIME CONFIGURATION FOR “PREP” AND COLLIMATOR LIGHT
Refer to Section 2.2.4
C05
C5.1
“Prep”
Time Configuration:
10 - 50 seconds
Factory Configured at 30 seconds.
C5.2
“Collimator Light”
Time Configuration:
10 - 50 seconds
Factory Configured at 30 seconds.
MAXIMUM X-RAY TUBE POWER
Refer to Section 2.2.5
C06
C6.1
Maximum percentage limit for X-Ray Tube
Select the maximum power percentage
at 100, 90, 80 or 70.
MISCELLANEOUS
Refer to Section 2.2.6
C7.1
Speaker Sound: OFF / ON
OFF: all sounds disabled except for Exp.
ON: all sounds enabled (factory set)
.
C07
C7.2
Winter / Summer Automatic Time Update:
NO / YES
NO: automatic time update disabled.
YES: automatic time update enabled
(factory set).
C7.3
Collimator Light ON with PREP: OFF / ON
OFF: Collimator Light only enable with button
(factory set).
ON: Collimator Light enable when pressing
“
PREP
”.
DATE AND TIME
Factory set at GMT +1.Refer to Section 2.2.7
C8.1
Hour
0 - 23
C08
C8.2
Minutes
0 - 59
C08
C8.3
Date
1 - 31
C8.4
Month
1 - 12
C8.5
Year
2000 - 2099
If any of the configurated values is not consistent with the Unit
Parameters, the system will not allow to exit from Configuration
mode. Double check the configured values (Refer to Table 2-1).
Note
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...