Portable X-ray Units
Service Manual
SM-1076R2
108
5.3.16.5 INDUCTANCE CHOPPER
1.
Access inside the Central Support as described in
Section 5.3.16.1
.
2.
Cut the Tie Wrap if needed.
3.
Cut the Heat Shrinks and unsolder the two (2) wires attached keeping in
mind the correspondence of each cable.
4.
Unscrew the four fixing screws located at the side of the Central Support.
5.
Dismount the Inductance Chopper from its support.
Fixing Screws
6.
Assemble the new Inductance Chopper on its support and then in the
Central Support.
7.
Solder the two (2) wires in the respective terminals covering both cables
with the new Heat Shrinks and applying heat to reduce the Heat Shrinks.
8.
Assemble the Unit as described in
Section 5.3.16.1
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...