Portable X-ray Units
Service Manual
SM-1076R2
62
ERROR CODE :
E37
DESCRIPTION
:
Tube Overload error. The Tube ratings are exceeded or not enough remaining
Thermal Capacity of the Tube to perform the selected exposure.
This error does not require to press the “
Reset
” push-button, the error code
disappears automatically.
POSIBLE CAUSES
The selected exposure is not possible due to the Tube thermal status.
The selected technique is beyond the X-ray Tube ratings or present conditions of the
X-ray Tube inhibit the exposure (anode overheated).
Parameters for next exposure may be temporally limited by the Unit.
ACTIONS
1.
Check that the remaining thermal capacity (%) is lower than the calculated for the
next exposure (thermal capacity close to 0%).
2.
Wait until the Tube cools down.
3.
If the error code does not disappear after 15 minutes, turn the Unit OFF and ON.
4.
If the equipment remains inoperative, check the hardware and replace if needed.
ERROR CODE :
E38
DESCRIPTION
:
System error -- Chopper failure.
The DC Bus voltage and / or the Chopper Coil current is/are too high.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
Over voltage or over current in the Chopper, noise or fault in the hardware that measure
these parameters.
ACTIONS
1.
Turn Off/ON the unit.
2.
If the error persists, replace the defective hardware.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...