Portable X-ray Units
Service Manual
SM-1076R2
41
Table 5-1
Messages in the Touch Screen Console
MESSAGE
DESCRIPTION
Messages related to Generator Power Limit:
“kVp RANGE LIMIT”
“mAs RANGE LIMIT”
“mA RANGE LIMIT”
“ms RANGE LIMIT”
If the power limit (kVp x mA) is reached by increasing the kVp or mAs
up to a maximum possible value, the mAs value is blocked.
Flashing values on kVp and mAs displays and an audible signal will
alert operator about the situation with one of these messages.
If required, kVp could be increased up to its maximum value while
mA value may automatically decrease, as long as mAs value is kept
the same.
“SPACE CHARGE LIMIT”
If a variation of the kVp or mAs induces to reach space charge limit,
the parameter is blocked, and a flashing value on the kVp display
and an acoustic signal will alert operator about the situation.
Messages related to X-ray Tube or Heating:
“OVERHEAT”
“TUBE OVERHEAT”
“INVERTER OVERHEAT”
“INVERTER LOAD LIMIT”
If a technique reaches the X-ray Tube ratings limit or the X-ray Tube
is momentarily overheated, some technique could not be selected.
Flashing values on the kVp / mAs displays and an acoustic signal
will alert operator about the situation.
“X-RAY KEY”
The X-Ray Key is not in on position.
This message only applies to Generators with the “X-Ray Key
option”.
“
TUBE POWER LIMIT”
The technique overpasses the Tube Power Limit.
“FOCUS CHANGE INHIBIT”
The selected technique does not allow the Focus change.
“LINE POWER TOO LOW”
The Power Line where the unit is connected will bear the usage of
the Unit.
“WRONG APR WORKSTATION”
The selected workstation is not allowed for the APR.
“WRONG APR TECHNIQUE”
The APR Technique selected is not allowed.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...