Portable X-ray Units
Service Manual
SM-1076R2
47
ERROR CODE :
E02
DESCRIPTION
:
Wrong Workstation configuration. One or more workstations are not properly configured,
a default value has been assigned. “Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The workstation configuration is invalid, for example a workstation configured as
bucky/digital panel (C1.2 or C2.2) has no input selected (C1.3 or C2.3). Or the EEPROM
is not initialized, that is very uncommon.
ACTIONS
1.
Review the workstation configuration (C01 and C02 menus) and correct the
wrong value.
2.
Consider initializing the EEPROM (
refer to Section 5.1.1
). If the EEPROM is
initialized, the calibration data will be lost and the system configuration will be
overwritten with the default values.
ERROR CODE :
E03
DESCRIPTION
:
No Workstation configured. All the workstations have no Tube configured, there is no
workstation available, a default value has been assigned. “Recoverable” with the “
Reset
”
push-button.
POSIBLE CAUSES
The workstation configuration is invalid, or the EEPROM is not initialized, that is very
uncommon.
ACTIONS
1.
Review the workstation configuration (C1.1 and C2.1 submenus) and configure
at least one of them with a value different from “0”.
2.
Consider initializing the EEPROM (
refer to Section 5.1.1
). If the EEPROM is
initialized, the calibration data will be lost and the system configuration will be
overwritten with the default values.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...