ERROR CODE :
E51
DESCRIPTION
:
Incorrect exposure time. The exposure time is above the maximum (very uncommon) or
it is so short that it is not reachable according to the present configuration (rise time and
kVp tail / high voltage capacity). “Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The exposure time is shorter than the added of the 25% of the rise time plus the kVp tail
from the selected kVp to the 75% of that value.
ACTIONS
Change the exposure parameters increasing the exposure time.
ERROR CODE :
E60
DESCRIPTION
:
The preprogrammed exposures for Autocalibration has finished without complete the
Autocalibration process. The number of exposures to autocalibrate a mA station have
run out. This error can be only shown in Service Mode during Autocalibration process.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
Tube instability, defective filament control hardware, noise or fault in the mA measure.
ACTIONS
If the error persists, replace the defective hardware.
ERROR CODE :
E95
DESCRIPTION
:
Power Line not good enough for operation. The Generator could not finish properly the
“Automatic Line Power Detection” procedure. No exposure will be allowed until the user
performs a successful test. “Unrecoverable”, the system remains lock.
POSIBLE CAUSES
The line impedance is too high.
ACTIONS
1.
Repeat the “Automatic Line Power Detection” procedure (
refer to the Operator
Manual
).
2.
If error persists, plug the unit to another socket line and try again.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...