Portable X-ray Units
Service Manual
SM-1076R2
121
5.
Unscrew the nuts located at each corner of the Portable Control Board
and remove it.
6.
Install the new Portable Control Board.
7.
Reconnect all the cables and assemble in reverse order.
8.
Initialize the new EEPROM (
refer to Section 5.1.1
).
5.4.8
REAL TIME CLOCK BATTERY
To replace the Real Time Clock Battery is not necessary to remove
the X-Ray Power Unit from the Trolley, it can be replaced in situ.
1.
Access to the Portable Control Board (
A3175--03
) as described in
Section 5.4.7 (steps 1. to 3.)
.
2.
Take the battery BAT1 (model CR 1216, 3V) out of the Portable Control
Board (
A3175--03
) and replace it.
3.
Assemble in reverse order.
Note
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...