SECTION 1
INSTALLATION REQUIREMENTS
The Portable Unit has been Configured, Calibrated and
Tested by the Manufacturer. It is ready for normal operation.
No service task is required. Refer to this Service Manual only
for Periodic Maintenance or repairs.
This Service Manual covers the Portable Units for Monoblock
and the Monoblock with Trolley versions.
1.1
POWER LINE REQUIREMENTS
The Unit is factory provided with a Power Cable (6 meters) and a
Plug for connection to line voltage (wall socket) according to
customer order. If required, change the line cord and plug to
conform to local codes and requirements.
ACCORDING TO MDD/93/42/CEE, THIS UNIT IS EQUIPPED
WITH EMC FILTERS. THE LACK OF PROPER GROUNDING
MAY PRODUCE ELECTRICAL SHOCK TO THE USER.
Portable X-ray Units power requirements are:
Maximum Power kW
2.0
4.0
8.0
Single-Phase, 100 -- 240 VAC, 50 / 60 Hz.
Single-Phase,
220 -- 240 VAC, 50 / 60 Hz.
Power Line Operation
Line voltage automatic compensation:
±
10%
Power line cable of the Portable Unit: 6 meters.
Connection to standard outlets with GND that accomplishes local regulations.
Maximum Power Line Impedance
Refer to Illustration 1-1
The General Circuit Breaker installed in the Portable Unit is 32 A (curve type C)
with a 30 mA Sensitivity Differential.
The Power Line Installation should be provided with a 30 mA Sensitivity Differential
and with a Thermomagnetic Interruptor / Circuit Breaker of at least:
Minimum recommended
Thermomagnetic / Circuit Breaker
²
16 A (curve type C) or
²
10 A (curve type D)
for 100 -- 120 VAC
²
8 A (curve type C) or
²
6 A (curve type D)
for 220 -- 240 VAC
²
30 A (curve type C) or
²
16 A (curve type D)
for 100 -- 120 VAC
²
16 A (curve type C) or
²
10 A (curve type D)
for 220 -- 240 VAC
²
30 A (curve type C) or
²
16 A (curve type D)
for 220 -- 240 VAC
Momentary Line Current based on 100 ms X-ray exposure (RMS)
Note
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...