Portable X-ray Units
Service Manual
SM-1076R2
28
5.
Make an exposure again to verify that the mA value in the Oscilloscope
is the one selected; if it is not, repeat the procedure from step 4.b. to
increase or decrease the value.
6.
Once the correct mA value is reached, save the value pressing
“Reset”
push-button (one beep will sound and the saved value is shown on the
kVp Display). It is important to write down this data for later reference
during calibration (
refer to Table 3-3
).
Once a value for Filament Current is saved, the value remains in
memory for the next mA station. This feature helps to save time at
entering new data for the next station.
7.
Repeat above procedure to calibrate the Filament Current for the
following kVp station and the non calibrated mA stations, in the
corresponding Focal Spot. The Filament Current value is few steps
above the one calibrated before (
refer to Table 3-3
).
Values above 999 are shown on mAs display as “= XX”, a.e. “=23“
means 1023.
8.
To exit Calibration Mode, press and hold
“Collimator Light”
push-button
and then press and release
“mAs Increase”
push-button (two beeps
sound). Then release
“Collimator Light”
push-button.
At the end of the Filament Current calibration process there are
some stations that do not admit exposure, in that cases enter
manually the value following the trend. (Refer to Table 3-2 as an
example).
KEEP IN MIND THAT ALL FILAMENT CURRENTS SHOULD
HAVE A VALUE, EVEN THE ONES OUT OF RANGE
(EXPOSURES NOT ALLOWED). IN THAT CASES A
CONTINUOUS BEEP WILL SOUND AND THE VALUES ON THE
DISPLAYS BLINK.
IF THE DIP-SWITCH SW1-1 OF THE PORTABLE CONTROL
BOARD HAS BEEN SET IN “ON” POSITION TO ENABLE
SERVICE MODE, DO NOT FORGET TO SET IT IN “OFF”
POSITION (SERVICE MODE DISABLED) ONCE THE
CALIBRATION IS FINISHED.
Note
.
Note
.
Note
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...