Portable X-ray Units
Service Manual
SM-1076R2
31
3.2.6
MANUAL MODIFICATION OF kVp GAIN AND mA GAIN
In some situations in which it is difficult to perform the calibration
of kVp Gain and/or mA Gain or when the value is already known,
these values can be manually set.
1.
To enter Calibration Menu,
press and hold
“Collimator Light”
push-button and then press and release
“mAs Decrease”
push-button
(two beeps sound). (CAL PArA) appears on the displays while
“Collimator Light”
is pressed
.
Then release
“Collimator Light”
push-button. The kVp Display shows P01 and one of the Focal Spot Leds
blinks.
1
2
Press and hold Collimator Light
Press and release mAs Decrease
Release Collimator Light
3
kVp
mAs
In case you have to manually modify the mA Gain instead of the
kVp Gain, select menu P02 by pressing the “kVp Increase”
push-button (kVp Display shows P02).
2.
Press any
“Focal Spot”
push-button to check the current Gain value (kVp
o mA Gain), it is displayed on the mAs display for a couple of seconds.
3.
Press
“mAs Decrease” or “mAs Increase”,
to set the new value.
4.
Press
“Reset”
push-button to save the value, a long beep sounds.
5.
Exit Calibration by pressing and holding “
Collimator Light”
push-button
and then press and release
“mAs Decrease”
(two beeps sound). Then
release
“Collimator Light”
(kVp and mAs parameters are shown).
IF THE DIP-SWITCH SW1-1 OF THE PORTABLE CONTROL
BOARD HAS BEEN SET IN “ON” POSITION TO ENABLE
SERVICE MODE, DO NOT FORGET TO SET IT IN “OFF”
POSITION (SERVICE MODE DISABLED) ONCE THE
CALIBRATION IS FINISHED.
Note
.
Note
.
Содержание SPL-HF 2.0
Страница 6: ...Portable X ray Units Service Manual...
Страница 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Страница 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Страница 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Страница 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Страница 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Страница 179: ...Portable X ray Units Service Manual SM 1076R2 169 PORTABLE STAND B1 B2 B3 B4 B6 B12 B7 B5 B9 B13 B8 B15 B9 B11 B10 B14...
Страница 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Страница 181: ...Portable X ray Units Service Manual SM 1076R2 171 7 2 MONOBLOCK POWER UNIT WITH TROLLEY A0 C0 D0 E0...
Страница 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Страница 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Страница 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...