Select Waveform Interpolation Method
Under real-time sampling, the oscilloscope acquires the discrete sample values of the
waveform being displayed. In general, a waveform of dots display type is very difficult to
observe. In order to increase the visibility of the signal, the digital oscilloscope usually
uses the interpolation method to display a waveform.
Interpolation method is a processing method to
―
connect all the sampling points
ǁ
, and
using some points to calculate the whole appearance of the waveform. For real-time
sampling interpolation method is used, even if the oscilloscope in a single captures only a
small number of sampling points. The oscilloscope can use interpolation method for filling
out the gaps between points, to reconstruct an accurate waveform.
Press the
Acquire
button on the front panel to enter the ACQUIRE Function menu; then
press the
Sinx/X
softkey to select
Sinx
or
X
.
X:
In the adjacent sample points are directly connected on a straight line. This
method is only confined to rebuild on the edge of signals, such as square wave.
Sinx:
Connecting
the
sampling
points
with
curves
has
stronger
versatility.
Sinx interpolation method uses mathematical processing to calculation
results in the actual sample interval. This method bending signal waveform, and
make it produce more realistic regular shape than pure square wave and pulse. When
the sampling rate is 3 to 5 times the bandwidth of the system. Recommended sinx
interpolation method.
Figure 11 Display Type Set to Dots
Содержание SMO1000A-S
Страница 1: ...Digital Oscilloscope SMO1000A S User Manual ...
Страница 50: ...User Manual SMO1000A S 50 176 Figure 12 x Interpolation Figure 13 Sinx Interpolation ...
Страница 76: ...User Manual SMO1000A S 76 176 Figure 28 Relative Window Trigger ...
Страница 78: ...User Manual SMO1000A S 78 176 Figure 29 Interval Trigger ...
Страница 82: ...User Manual SMO1000A S 82 176 Figure 32 Runt Trigger ...
Страница 96: ...User Manual SMO1000A S 96 176 Figure 44 SPI Trigger ...