User Manual SMO1000A-S
115/176
To Make Cursors Measurements
Cursors are horizontal and vertical markers that indicate X- axis values and Y- axis values
on a selected waveform source. You can use cursors to make custom voltage, time
measurements on oscilloscope signals.
X Cursors
X cursors are vertical dashed lines that adjust horizontally and can be used to measure time
(when the source is FFT waveform, X cursors measure frequency)
X1
cursor is the left (default position) vertical dotted line; it can be moved to any place of the
screen.
X2
cursor is the right (default position) vertical dotted line; it can be moved to any place of
the screen.
Use the
Universal Knob
to set the
X1
and
X2
cursor values and the values are displayed in
the cursors box in the upper-left corner of the screen along with the difference between
X1 and X2 (
△
T) and 1/
△
T.
When set cursor type to
X1-X2
, use
Universal Knob
will move the X1 and X2 cursors
together. The value under the menu option is the difference between the X1 and X2 cursors.
Y Cursors
Y cursors are horizontal dotted lines that adjust vertically and can be used to measure
voltage (V) or current (A). When the cursors source is the math function, the unit will match
the math function.
Y1
cursor is the top (default position) horizontal dotted line; it can be moved to any vertical
place of the screen.
Y2
cursor is the down (default position) horizontal dotted line; it can be moved to any vertical
place of the screen.
Use the
Universal Knob
to set the Y1 and Y2 cursor values and the values are displayed in
the cursors box in the top left corner of the screen along with the difference between Y1
and
Y2 (
△
Y).
When set cursor type to
Y1-Y2
, use
Universal Knob
will move the Y1 and Y2 cursors
together. The value under the menu option is the difference between the Y1 and Y2 cursors.
Содержание SMO1000A-S
Страница 1: ...Digital Oscilloscope SMO1000A S User Manual ...
Страница 50: ...User Manual SMO1000A S 50 176 Figure 12 x Interpolation Figure 13 Sinx Interpolation ...
Страница 76: ...User Manual SMO1000A S 76 176 Figure 28 Relative Window Trigger ...
Страница 78: ...User Manual SMO1000A S 78 176 Figure 29 Interval Trigger ...
Страница 82: ...User Manual SMO1000A S 82 176 Figure 32 Runt Trigger ...
Страница 96: ...User Manual SMO1000A S 96 176 Figure 44 SPI Trigger ...