7- bit Address & Data Condition
—The oscilloscope triggers on a read or write
frame in 7- bit addressing mode on the 17th or the 26th clock edge if all bits in the
pattern match. For triggering purposes, a restart is treated as a start condition
On the 17th Clock Edge
On the 26th Clock Edge
10- bit Address & Data Condition
— The oscilloscope triggers on a 10- bit write
frame on the 26th (or 34th) clock edge if all bits in the pattern match. The frame is
in the format:
Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data) The oscilloscope
triggers on a read or write frame in 10- bit addressing mode on the 26th clock
edge if all bits in the pattern match.
Data 2
shows
oxXX
. For triggering purposes,
a restart is treated as a start condition.
Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data: Ack: Data) The
oscilloscope triggers on a read or write frame in 7- bit addressing mode on the
34th clock edge if all bits in the pattern match.
Data2
is available. For triggering
purposes, a restart is treated as a start condition