User Manual SMO1000A-S
118/176
Type of Measurement
Voltage Measurements
Voltage measurements include 15 kinds of voltage parameter measurements.
Figure 60 Voltage Measurements
1.
Vpp:
The peak- to- peak value is the difference between Maximum and Minimum
values.
2.
Vmax:
Maximum is the highest value in the waveform display.
3.
Vmin:
Minimum is the lowest value in the waveform display.
4.
Vamp:
The Amplitude of a waveform is the difference between its Top and Base values.
5.
Vtop:
The Top of a waveform is the mode (most common value) of the upper part of
the waveform, or if the mode is not well defined, the top is the same as Maximum.
6.
Vbase:
The Base of a waveform is the mode (most common value) of the lower part of
the waveform, or if the mode is not well defined, the base is the same as Minimum.
The Y cursor shows the value being measured.
7.
Vmea:
The arithmetic mean over the first cycle of the waveform.
8.
Mean:
The arithmetic over the entire waveform.
9.
Stdev:
Standard deviation of all data values
10. Vstd:
Standard deviation of all data values in the first cycle
11. Vrms:
The true Root Mean Square voltage over the entire waveform.
12. Crms:
The true Root Mean Square voltage of the first cycle in the waveform.
13. Overshoot:
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. ROV means rising edge overshoot and FOV means falling
edge overshoot.
Содержание SMO1000A-S
Страница 1: ...Digital Oscilloscope SMO1000A S User Manual ...
Страница 50: ...User Manual SMO1000A S 50 176 Figure 12 x Interpolation Figure 13 Sinx Interpolation ...
Страница 76: ...User Manual SMO1000A S 76 176 Figure 28 Relative Window Trigger ...
Страница 78: ...User Manual SMO1000A S 78 176 Figure 29 Interval Trigger ...
Страница 82: ...User Manual SMO1000A S 82 176 Figure 32 Runt Trigger ...
Страница 96: ...User Manual SMO1000A S 96 176 Figure 44 SPI Trigger ...