User Manual SMO1000A-S
67/176
Slope Trigger
The slope trigger looks for a rising or falling transition from one level to another level in
greater than or less than a certain amount of time.
In the oscilloscope, positive slope time is defined as the time difference between the two
crossing points of trigger level line A and B with the positive edge as shown in the figure
below.
1.
Press the
Setup
button on the front panel to enter the TRIGGER function menu.
2.
Press the
Type
softkey; turn the
Universal Knob
to set select
Slop
and then push the
knob to confirm.
3.
Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as the
trigger source.
4.
Press the Slop softkey; turn the
Universal Knob
to set select the desired trigger edge
(rising or falling), and then push down the knob to confirm. The current trigger slope is
displayed at the upper right corner of the screen.
5.
Press
Lower Upper
softkey to select the
Lower
or
Upper
trigger level; then turn the
Trigger Level Knob
to adjust the position. The trigger level values are displayed at
the upper right corner of the screen.
The lower trigger level cannot be upper than the upper trigger level. In the trigger state
message box, L1 means the upper trigger lever while L2 means the lower trigger level.
Содержание SMO1000A-S
Страница 1: ...Digital Oscilloscope SMO1000A S User Manual ...
Страница 50: ...User Manual SMO1000A S 50 176 Figure 12 x Interpolation Figure 13 Sinx Interpolation ...
Страница 76: ...User Manual SMO1000A S 76 176 Figure 28 Relative Window Trigger ...
Страница 78: ...User Manual SMO1000A S 78 176 Figure 29 Interval Trigger ...
Страница 82: ...User Manual SMO1000A S 82 176 Figure 32 Runt Trigger ...
Страница 96: ...User Manual SMO1000A S 96 176 Figure 44 SPI Trigger ...