User Manual SMO1000A-S
167/176
AWG Self Cal
If the operating temperature changes more than 5°and the instrument runs more than
more 30 minutes, you should do the AWG Self Cal to decrease the mistake.
1.
Press the
Wave Gen
button on the front panel to enter the WAVEFORM function
menu.
2.
Press the
Setting
softkey to enter the SETTING function menu.
3.
Press the
AWG Self Cal
softkey to perform AWG self calibration program.
Figure 92 AWG Self Cal
4.
When the self calibration program is finish, press the
Run/Stop
button on the front
panel to exit the interface according to the information on the screen.