RM-804 & RM-805 Instruction Manual/English
96
13/10/2016 Version No. 001
20
Relay Controls the relay output.
Out
22
STRB After all (up to 100) output groups are ready, the
STRB signal will pulse high.
Out
Scan Setup
Background
The Scan function sequentially scans up to 100 channels and
grades the resistance of the DUT on each channel to a reference
value. An automated handler or test fixture is required to interface
the DUTs to the measurement terminals and the scan interface
that controls the timing of each scan.
RM-805
Scan I/O
Measurement terminals
Automated Handler/Test Fixture
1
2
3
4
5
6
N
Channel
DUT 1~N
S+
Source- S-
·····
Note
The automated handler/test fixture is user-supplied. Please see
your distributor for support and technical details.