RM-804 & RM-805 Instruction Manual/English
36
13/10/2016 Version No. 001
This produces a positive and negative measurement voltage
across the DUT, which also includes the Vemf (V1+Vemf &
V2+Vemf).
0V
t
Vemf
V1
V2
To cancel the Vemf, V2 is deducted from V1 and divided by 2 to
get the average measurement, as shown in the formula below:
2
)
2
(
)
1
(
Vemf
V
Vemf
V
Vx
Where Vx = measured voltage sans Vemf.
Select Measuring Signal (Drive)
Background
Resistance measurement has 5 different measuring signals that
can be applied to obtain a resistance measurement: DC+, DC-,
Pulse, PWM, Zero.
Note
The Dive function is only applicable to the RM-805. The drive
signal for the RM-804 is fixed to DC+.