RM-804 & RM-805 Instruction Manual/English
174
13/10/2016 Version No. 001
TEMPerature:CONVersion:RESistance
Set
Query
Description
Sets or returns the initial resistance for the temperature
conversion function.
Syntax
Query Syntax
TEMPerature:CONVersion:RESistance {<NRf>[,<String>]}
TEMPerature:CONVersion:RESistance?
Parameter
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter
<NR3>
000.0001~999.9999E±X
Example
TEMP:CONV:RES 10.00,maohm
Sets initial resistance value to 10.00MΩ.
TEMP:CONV:RES?
>10.0000E+6
Returns the initial resistance as 10.00MΩ.
TEMPerature:CONVersion:TEMPerature
Set
Query
Description
Sets or returns the initial temperature for the temperature
conversion function.